What to do after a failed spore test
Also called: positive spore test, positive BI
Take the sterilizer out of service, check procedures, and retest right away.
A positive (failed) spore test means spores survived. CDC's steps: take the sterilizer out of service, review loading, packaging and monitoring records with everyone who uses it, recall and quarantine any implant items, and repeat the spore test right away using the same cycle. If the repeat test passes and the other indicators look right, it can go back into service. If it fails again, the sterilizer stays out of service until repaired and passes spore tests in three cycles in a row, and items back to the last negative test should be recalled and reprocessed when possible.
What you do
Safety and care
- Common causes are poor cleaning, running the sterilizer incorrectly, improper loading or overloading, and improper or wrong-type packaging.